Chapter 3: Setting up the BenchCel Workstation
BenchCel Microplate Handling Workstation R-Series User Guide
68
3.
In the
Device List
area, select a BenchCel device and click
Device
diagnostics.
The BenchCel Diagnostics dialog box opens.
4.
In the
Profile
tab, check the
Teachpoint File
area to make sure the
correct teachpoint file is loaded. You will be adding new
teachpoints to this file.
!! IMPORTANT !!
If you have not already done so, make a
backup copy of this file.
5.
In the
Controls
tab, select the desired labware definition from the
Labware
list that is below the graphical display area.
Check the teachpoint file path.
Select labware.