113
Chapter 5: Maintenance and troubleshooting
BenchCel Microplate Handling Workstation R-Series User Guide
Opening and closing
clamps
The clamps in the stacker head close and open the grippers at the
bottom of the stacker rack to hold and release the first plate in position
for the robot grippers. Compressed air is used to move the clamps.
The clamps close and open the stacker grippers automatically during
the loading, unloading, downstacking, and stacking procedures. When
diagnosing problems or after an aborted run, you can use the
commands in BenchCel Diagnostics to open or close the clamps
(stacker grippers). For example, you might want to open the clamps
(stacker grippers) to remove a microplate.
!! DAMAGE HAZARD !!
Opening the clamps (stacker grippers)
might cause the microplate or stack of microplates to drop.
To open or close the clamps (stacker grippers):
In the
BenchCel Diagnostics Controls
tab, click
Stacker n
at the top of the
rack that you want to load or unload. In the command menu that
appears, click
Open Stacker Grippers
or
Close Stacker Grippers.
You
should hear a click sound as the clamps (stacker grippers) open or
close.
The clamps open or
close the stacker
grippers.
Click the command at the desired rack to
open or close the stacker grippers (clamps).
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...