5
BenchCel Microplate Handling Workstation User Guide Addendum
9. In the
Jog/Teach
tab, type the calculated
Plate presence threshold.
Determining the
optimum Orientation
sensor offset
To determine the optimum Orientation sensor offset:
1. In BenchCel Diagnostics, click
Downstack
at the first stack.
2. In the
Controls
tab, click the
Labware
tab.
3. In the
Orientation sensor offset
box, type the initial offset value you
want to use. To determine the initial value:
a. Determine the halfway distance (in millimeters) between the
top of the microplate and the top of the skirt.
b. Determine the height of the microplate skirt (in millimeters).
c.
Add the values from step a and step b.
Divide this height by 2.
Add the skirt height.