Chapter 5: Maintenance and troubleshooting
BenchCel Microplate Handling Workstation R-Series User Guide
120
9.
To make changes to other labware parameters, click
Editor.
You can
make changes in the Labware Editor dialog box. See the
BenchWorks Automation Control User Guide
for instructions on using
the Labware Editor.
Lid departure
height
The height, in millimeters, above the bottom of the
plate to which the lid is lifted.
Gripper holding lid
position
The distance, in millimeters, that each gripper
moves inward from its home position when
holding a microplate lid. An increasing value
moves the grippers closer together and holds the
microplate tighter. A decreasing value opens the
grippers wider.
The parameter value is applied to each robot
gripper. For example, a value of 5.25 mm moves
each robot gripper 2.625 mm toward each other
from its home position.
In general, type a value that is less than Gripper
holding plate position to open the grippers slightly.
Holding the lid too tightly might cause the
microplate to be lifted with the lid.
Parameter
Description
c
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...