127
Chapter 5: Maintenance and troubleshooting
BenchCel Microplate Handling Workstation R-Series User Guide
6.
In the graphical display area, at the stack where you have
downstacked the first microplate, click
Move to Sensor.
The robot
moves the microplate into the line of sight of the plate-orientation
sensors.
7.
In the
Jog/Teach
tab, in the
Stacker Sensors
area, select the stacker
number, and then check the notch sensor readings. The corners
with the notches should have much lower readings than the other
corners.
In the following example, the plate notch is in the upper right
corner, so the reading at that corner is significantly lower than the
other corners.
8.
If the corners with notches do not have lower readings than the
other corners, in the
BenchCel Diagnostics Controls
tab, click the
Labware
tab.
Click to move the plate to the
plate-orientation sensors.
Select the stacker number.
Check the sensor readings.
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...