163
Appendix B: Quick reference
BenchCel User Guide
Plate Dimensions area
Parameters
Description
Thickness
The distance, in millimeters, from the bottom
surface of the plate to the top surface of the
plate.
For a tip box, this is the distance from the
bottom surface of the box to the top of the
tips.
To increase the number of contact points,
measure the distance at the corner of the
plate or tip box (using calipers). This method
is especially useful if the plate has a lip at the
top and the caliper can angle inward,
producing inaccurate measurements.
Stacking thickness
The thickness, in millimeters, of two stacked
plates minus the thickness of one plate.
For example:
Thickness of two stacked plates
(
x
) = 23.14 mm
Thickness of one plate = 14.14 mm
Stacking thickness: 23.14 mm - 14.14 mm =
9.00 mm
Vstack
plate and stacking thickness
00022
Plate Thickness
Stacking Thickness
x
00022
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...