Appendix B: Quick reference
BenchCel User Guide
168
BenchCel Diagnostics - General Settings tab
Parameters
Description
Plate presence threshold
Any value at or higher than this means a
microplate is present. Below this value means
there is no microplate.
Rack sensor threshold
Any value at or higher than this means a rack
is present. Below this value means there is no
rack.
Low pressure threshold (psi)
The value at or below which the system will
display a low-pressure warning.
Note: If you set this value too low, the device
could drop microplates without generating a
low air-pressure warning. Velocity11
recommends that you set the threshold at
20 psi.
Home robot after protocol
run finishes
The option to return the robot to its home
position after the protocol run is finished.
Speed settings
Opens the Speed Settings dialog box and
allows you to set each speed (Slow, Medium,
and Fast) as a percentage of the factory-set
maximum speed.
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...