11
BenchCel Microplate Handling Workstation User Guide Addendum
Worksheet for setting sensor thresholds
Plate-presence
threshold
Initial Plate present readings
Set the thresholds as follows:
❑
Plate presence threshold = 225 units
❑
Intensity = 100%,
Record the
Plate present
reading for each stack.
Plate present readings when microplates are unloaded
After adjusting the Intensity so that the highest Plate present reading is
less than or equal to 175, record the
Plate present
readings for each
stack.
Highest reading among all the stacks (P
unloaded
): _________________
Plate present readings when the microplates are loaded
After loading the microplates at each stack (
Load All
in BenchCel
Diagnostics), record the
Plate present
readings for each stack.
Lowest reading among all the stacks (P
loaded
): _________________
Plate presence threshold
Plate presence threshold = P
unloaded
+ ((P
loaded
– P
unloaded
)/2) = ______
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...