4
Maintenance
Barcode errors
Seahorse XFe Analyzer Operating Manual
35
Cell plate barcode read failure
For any Cell Plate barcode read errors, Wave Controller displays a message and
two corrective actions. (See
•
Manual
: Manually input the Cell Plate barcode info.
•
Cancel Assay
: Cancel the assay.
Manually enter cell plate barcode information as follows:
1
Click the Tray Widget. Cartridge Barcode Read Failure dialog box appears.
(See
2
To eject the Cell Plate, click
Open Tray
.
3
The Cell Plate barcode is located on the side of the plate. Write down the
barcode information.
4
Click
Close Tray
. The Cartridge Barcode Manual Entry window appears.
(See
5
Enter the Cell Plate barcode, and click
Accept
.
Figure 15.
Cell Plate Barcode Read Failure dialog box
Summary of Contents for Seahorse XFe
Page 1: ...Seahorse XFe Analyzer Operating Manual...
Page 4: ...4 Seahorse XFe Analyzer Operating Manual...
Page 12: ...1 Introduction Technical Specifications 12 Seahorse XFe Analyzer Operating Manual...
Page 38: ...4 Maintenance Additional Resources 38 Seahorse XFe Analyzer Operating Manual...
Page 39: ......