SureScan System User Guide
64
9
Turn off the power to the scanner.
10
Wait 10 seconds, and then turn on the scanner power.
11
Start the Scan Control program.
The Scan Control program and scanner firmware are now updated. If you have problems,
contact Agilent technical support.
To verify the software installation
The SureScan Microarray Scanner System workstation includes an installation qualification tool
(IQT). Use this program after updating your software to verify that the update installed correctly.
1
Click
Start > All Programs > Agilent Technologies > Installation Qualification Tool
.
Figure 16
Agilent Installation Qualification Tool dialog box
2
Mark the check box next to
Agilent Microarray Scan Control
.
3
Under
Products to be qualified
, click
Qualify
.
The installation is verified, and an installation qualification report is generated. If you selected
Open report after qualification
, the installation qualification report opens in your internet
browser.
4
When finished, click the close button in the upper right corner of the program dialog box.