TriScroll 620 Series Vacuum Pump 

8

Installation Requirements

Do not install or use the pump in an environment
exposed to atmospheric agents (rain, snow, ice), dust,
aggressive gases, or in explosive environments or those
with a high fire risk. 

During operation, the following environmental
conditions must be respected:

Operation

In order to reach ultimate vacuum, the pump must be left
running for about an hour with the inlet sealed.

Unlike conventional oil-sealed pumps, Agilent’s dry scroll
pumps do not have fluid to cleanse them of accumulated
dust and debris. Run the pump periodically at atmosphere
for a minute or two to flush out the pump. Flush the pump
regularly and adjust this schedule according to your
specific conditions.

Start up Procedure

1. Be sure that the vacuum system isolation valve is 

closed.

2. Turn on power to the pump.

3. Open the isolation valve.

Shutdown Procedure

1. Close the vacuum system isolation valve. This 

prevents debris in pump from being transported into 
the vacuum system.

2. Turn off power to the pump.

Maintenance

Personnel responsible for pump operation and
maintenance must be well-trained and aware of the
accident prevention rules.

If a pump is to be discarded, it must be disposed of in
accordance with specific national and local standards.

Temperature:

+5 °C to +40 °C (41 °F to 104 °F)

Relative humidity:

0 to 95% (non-condensing)

CAUTION

As supplied from the factory, the pump 
is configured for low voltage. Verify that 
the configuration matches the supply 
voltage. 

If voltage changeover is required, 
configure the voltage as described in 
“Electrical Connections”  on page 13.

WARNING

During operation the outer surface of 
the motor housing can become hot. 
Avoid touching the motor housing 
during pump operation.

WARNING

The pump is designed for operation 
with neutral or noncorrosive fluids. It 
is absolutely forbidden to use it with 
potentially explosive, inflammable or 
poisonous substances.

WARNING

Death may result from contact with 

high voltages. Always take extreme care 
and observe the accident prevention 
regulations in force.

When the machine is powered up, 

be careful of moving parts and high 
voltages. 

If you have to perform maintenance 

on the pump after a considerable time 
in operation, allow the pump to cool as 
the temperature of the outer surface 
may be in excess of 55 °C (131 °F).

Always disconnect your power 

supply to the pump before beginning 
maintenance work.

NOTE

Before returning the pump to the factory 
for repair, the “Health and Safety” sheet 
attached to this instruction manual must 
be completed and sent to the local sales 
office. A copy of the sheet must be 
inserted in the pump package before 
shipping.

Summary of Contents for TriScroll 620 Series

Page 1: ...INSTALLATION AND OPERATION MANUAL Vacuum Products Division DRAFT 10 30 14 TriScrollTM 620 Series Dry Scroll Vacuum Pump Manual No 699904343 Revision G November 2014...

Page 2: ...TriScroll 620 Series Vacuum Pump Copyright 2014 Agilent Inc DRAFT 10 30 14 TriScroll 620 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Inc...

Page 3: ...earing Purge 16 Purge Kit 16 Troubleshooting 17 Maintenance 18 General Information 18 Related TriScroll Manuals 18 Maintenance and Tooling Kits 18 Factory Service Options 19 Accessories 19 Contacting...

Page 4: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Page 5: ...ng representatives of Seller which are inconsistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller Warra...

Page 6: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Page 7: ...fety protocol Storage When transporting and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective p...

Page 8: ...d of in accordance with specific national and local standards Temperature 5 C to 40 C 41 F to 104 F Relative humidity 0 to 95 non condensing CAUTION As supplied from the factory the pump is configured...

Page 9: ...Female 3 8 National Pipe Thread NW25 adapter provided Gas ballast Female 1 4 National Pipe Thread 40 Micron sintered filter provided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage te...

Page 10: ...ng with Dimensions Figure 2 Pumping Speed Curves 355 13 98 13 50 22 85 Intake NW25 NW25 Exhaust extension included but not installed Exhaust 3 8 NPT port swivels 360 300 11 80 472 18 57 Single Phase 1...

Page 11: ...Screws 3 13 Motor Electrical Cover Unpacking and Inspection The shipping container is a double carton 1 After opening the outer box remove the foam packing 2 Slit open the inner box 3 Lift the pump wi...

Page 12: ...the proper capacity The circuit breaker must be suitably located and easily reached The circuit breaker must be marked at the disconnecting device for the equipment with the on off symbol I O and meet...

Page 13: ...available to strain relieve the electrical supply cable The cable can be mechanically held under the electrical cover in the groove provided The groove is sized for 14 gage cable A 1 2 14 NPSM hole i...

Page 14: ...s in either direction and realign the two 2 locating pins on the scroll module with the mating frame holes Ensure that the rubber spider is still on the motor coupling and that the coupling teeth are...

Page 15: ...n it is stopped It will also close when the TriScroll motor thermal overload protection switch shuts down the pump Wiring Using the data in Table 4 verify that the VPI Valve chosen is compatible with...

Page 16: ...allast port See Purge Kit below Bearing Purge The bearing purge port item 6 on Figure 3 on page 11 is sealed on the TriScroll 620 and must not be opened or used Purge Kit A purge kit Agilent part numb...

Page 17: ...gth of power supply cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak Locate and repair leak Water in pump Flush pump with air or dry nitrogen Gas ballast plugged Replac...

Page 18: ...erform maintenance on TriScroll pumps is provided in kit form A description of each kit and ordering information is provided in Table 7 Table 6 Other Related Manuals Title Applicable TriScroll Model P...

Page 19: ...ntact Agilent Customer Service at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web site at http www chem agilent com en US Products Instrum...

Page 20: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank DRAFT 10 30 14...

Page 21: ...re product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product p pleas...

Page 22: ...or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled ou...

Page 23: ...ther Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Produc...

Page 24: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...