Chapter 6
137
Calibration and Compensation
Calibration and Compensation Procedure without Using Dedicated Test
Fixture
6. Calibration
and
Comp
ensatio
n
4. How to select the table range for executing the calibration
Select the table range for executing the calibration by making the appropriate setting in the
Select Calibration Table
area of the Calibration Dialog box window.
If you select
Active table only
, compensation will only be executed for the measurement
point table (active table) selected in “2. How to select the measurement point table” on
page 135.
If you select
Calibrate all tables
, compensation will be executed for all measurement point
tables.
5. How to measure open calibration data
Measure the open calibration data as follows:
Step 1.
Connect the open standard (0 S) to the coaxial terminal on the tip of the test head as a
calibration reference plane.
Step 2.
Click the
OPEN
button in the calibration window and measure the open calibration data.
Step 3.
Remove the open standard from the test head coaxial terminal.
6. How to measure short calibration data
Measure the short calibration data as follows:
Step 1.
Connect the short standard (0
Ω
) to the coaxial terminal on the tip of the test head as a
calibration reference plane.
Step 2.
Click the
SHORT
button in the calibration window and measure the short calibration data.
Step 3.
Remove the short standard from the test head coaxial terminal.
7. How to measure load calibration data
Measure the load calibration data as follows:
Step 1.
Connect the load standard (50
Ω
) to the coaxial terminal on the tip of the test head as a
calibration reference plane.
Step 2.
Click the
LOAD
button in the calibration window and measure the load calibration data.
Step 3.
Remove the load standard from the test head coaxial terminal.
8. How to measure DC open/short/load calibration data
When you want to use the dedicated open/short/load standard to measure DC resistance
(Rdc), follow these steps to execute additional DC open/short/load calibration.
Step 1.
Connect the DC open standard to the coaxial terminal on the tip of the test head as a
calibration reference plane.
Step 2.
Click the
DC OPEN
button in the calibration window and measure the DC open calibration
data.
Step 3.
Remove the DC open standard and connect the DC short standard.
Step 4.
Click the
DC SHORT
button in the calibration window and measure the DC short
calibration data.
Summary of Contents for Twelfth 4287A RF LCR
Page 8: ...8...
Page 36: ...36 Chapter2 Installation Guide Preparation for Power Supply Figure 2 17 Power cable options...
Page 114: ...114 Chapter5 Setting Measurement Conditions Setting up dc resistance Rdc measurement...
Page 250: ...250 Chapter11 Setting and Using the Control and Management Facilities System Recovery...
Page 294: ...294 AppendixA Manual Changes Manual Changes...
Page 328: ...328 AppendixF Error Messages Undefined header...