142
Chapter 6
Calibration and Compensation
Calibration and Compensation Procedure without Using Dedicated Test
Fixture
2. DUT connection terminal used as calibration reference plane
When you want to use the DUT connection terminal as a calibration reference plane,
connect the working standard to the DUT connection terminal to perform open/short/load
calibration (Table 6-6). No other calibration compensation is necessary.
1. How to check and change calibration kit definition
Check the definition of the working standard to be used and change it as required as
described in “How to Define Calibration Kit” on page 145.
2. How to select the measurement point table
To perform calibration data measurement for only one of the eight measurement point
tables, follow these steps to select the active table.
NOTE
When performing calibration data measurement for all of the measurement point tables at
once, it is not necessary to select individual measurement point tables.
Step 1.
Press the
[Stml Select]
key on the front panel to display the Stml Select softkey menu.
Step 2.
Press the
TABLE No. [ ]
key in the softkey menu to display the TABLE No. [ ] softkey
menu.
Step 3.
In the softkey menu, press the key (
TABLE 1
to
TABLE 8
) corresponding to the
measurement point table for which calibration data measurement should be performed.
3. How to display the Calibration Dialog box
Perform the following steps to display the Calibration Dialog box.
Step 1.
Press the
[Cal/Compen]
key on the front panel to display the Cal/Compen softkey menu.
Step 2.
Press the
CAL DIALOG
key in the softkey menu to display the Calibration Dialog box
Table 6-6
Calibration compensation to use DUT connection terminal as calibration
reference plane
Required calibration
Position to be calibrated
Standard used
Open/short/load calibration
*1
DUT terminal (as calibration
reference plane)
Working standard
*2
(open/short/load)
*1. Perform DC open/short/load calibration as required.
*2. Use the Agilent 4287A option 004 (working standard set). Open calibration and short calibration can be made by
opening or shorting between DUTs; it is not necessary to use a special standard that has the same shape as the
DUT.
Summary of Contents for Twelfth 4287A RF LCR
Page 8: ...8...
Page 36: ...36 Chapter2 Installation Guide Preparation for Power Supply Figure 2 17 Power cable options...
Page 114: ...114 Chapter5 Setting Measurement Conditions Setting up dc resistance Rdc measurement...
Page 250: ...250 Chapter11 Setting and Using the Control and Management Facilities System Recovery...
Page 294: ...294 AppendixA Manual Changes Manual Changes...
Page 328: ...328 AppendixF Error Messages Undefined header...