152
Chapter 6
Calibration and Compensation
Preventing Mistakes During Execution of Calibration and Compensation
Figure 6-24
DC resistance (Rdc) limit check turned ON
Step 5.
In the
Short-Load
box in the
Rdc Limit (Calibration)
area of the
Calibration/Compensation Wizard Setting dialog box, enter the limit values for DC
resistance measurement for short calibration and load calibration.
Step 6.
In the
Load-Open
box in the
Rdc Limit (Calibration)
area of the
Calibration/Compensation Wizard Setting dialog box, enter the limit values for DC
resistance measurement for load calibration and open calibration.
Step 7.
In the
Short-Open
box in the
Rdc Limit (Compensation)
area of the
Calibration/Compensation Wizard Setting dialog box, enter the limit values for DC
resistance measurement for short compensation and open compensation.
Step 8.
Press the
OK
button in the Calibration/Compensation Wizard Setting dialog box to end the
calibration and compensation wizard setting.
NOTE
The three Rdc limit values set as examples in Figure 6-24 are the setting values when 50-
Ω
load is used. For guidelines on determining the three Rdc limit values, see the chapter
“Preparations for Accurate Measurements” in the
Programming Manual
.
2. How to execute calibration and compensation
Step 1.
Press the
CAL WIZARD
key in the Cal/Compen softkey menu to start the calibration and
compensation wizard.
Step 2.
Execute calibration and compensation as prompted by the instructions on the screen.
When calibration and compensation data measurement is executed, the respective DC
resistance measurement values are shown in the wizard screen (lower part). If the Rdc does
not exceed the limit set in “1. Setting the limit for DC resistance (Rdc)” on page 151, a blue
mark (
) appears to the left of the DC resistance measurement value (Figure 6-25).
Summary of Contents for Twelfth 4287A RF LCR
Page 8: ...8...
Page 36: ...36 Chapter2 Installation Guide Preparation for Power Supply Figure 2 17 Power cable options...
Page 114: ...114 Chapter5 Setting Measurement Conditions Setting up dc resistance Rdc measurement...
Page 250: ...250 Chapter11 Setting and Using the Control and Management Facilities System Recovery...
Page 294: ...294 AppendixA Manual Changes Manual Changes...
Page 328: ...328 AppendixF Error Messages Undefined header...