Chapter 8
Sorting According to Measurement Results
Setting Upper and Lower Limit Values for DC Resistance Measurement
Setting Upper and Lower Limit Values for DC Resistance
Measurement
When the DC resistance (Rdc) measurement is activated, the 4287A starts DC resistance
measurement before impedance measurement. In this case, pass/fail evaluation is made
according to the preset test limit after DC resistance measurement. If the result deviates
from the test limit, the /RDC_OUT_OF_RANGE signal of the handler interface is
activated and the DC resistance measurement result is displayed in red.
Use the following steps to set the upper and lower limit values for the DC resistance
measurement.
Step 1.
Click with the mouse in the DC resistance upper limit value cell (
UP x.xx
Ω
) in the
upper-left corner of the BIN sort setup table.
Step 2.
Press the entry block numeric keys (
[0]
to
[9][-][.]
) to enter the upper limit value of the DC
resistance and then press the unit key (
[G/n]
,
[M/
μ
]
,
[k/m]
, or
[
×
1]
).
To set the upper limit value to 20
Ω
, for example, press
[2],
[0]
, and then
[
×
1]
Figure 8-14
Setting upper limit value for DC resistance (Rdc)
Step 3.
Set the DC resistance lower limit value by clicking in its cell and then following the
procedure of step 2.
Summary of Contents for Twelfth 4287A RF LCR
Page 8: ...8...
Page 36: ...36 Chapter2 Installation Guide Preparation for Power Supply Figure 2 17 Power cable options...
Page 114: ...114 Chapter5 Setting Measurement Conditions Setting up dc resistance Rdc measurement...
Page 250: ...250 Chapter11 Setting and Using the Control and Management Facilities System Recovery...
Page 294: ...294 AppendixA Manual Changes Manual Changes...
Page 328: ...328 AppendixF Error Messages Undefined header...