Chapter 8
193
Sorting According to Measurement Results
Example of Settings for BIN Sorting
8. Sorting According
to
M
e
as
ur
em
e
n
t Re
su
lt
s
Setting the measurement point number, measurement parameter, and
BIN range setting mode for Condition 1
Step 1.
Press the
[Cmptr]
key in the instrument state block (INSTR STATE) on the front panel to
display the BIN sorting setup display (Figure 8-17).
Figure 8-17
BIN sorting setup display
Step 2.
Use the following procedure to set the measurement point number for Condition 1.
a.
Click with the mouse the measurement point number (Stimulus) cell in the Condition 1
column to select the cell.
b.
Click the down-arrow
that appears in the right side of the cell.
c.
In the drop-down list box that appears, click “(3) 100.00 MHz” (Figure 8-18).