86
Chapter 4
Function Overview
Names and Functions of Display Areas on the LCD Display
NOTE
To directly call up the FREQUENCY softkey menu, click in the measurement frequency
area with the mouse. You can use this softkey menu to select the measurement point by the
measurement frequency instead of the measurement point number.
7. Oscillator level (power)
Displays the oscillator level of the currently selected measurement point. Use the
“Measurement point setup display” on page 91 to set the measurement point conditions
(measurement frequency, averaging factor, oscillator level for the measurement point
number) for each table.
NOTE
To directly call up the POWER softkey menu, click in the oscillator level area with the
mouse. You can use this softkey menu to select the measurement point by the oscillator
level instead of the measurement point number.
8. Averaging factor
Displays the measurement averaging factor of the currently selected measurement point.
Use the “Measurement point setup display” on page 91 to set the measurement point
conditions (measurement frequency, averaging factor, oscillator level for the measurement
point number) for each table.
NOTE
To directly call up the AVERAGE softkey menu, click in the averaging factor area with the
mouse. You can use this softkey menu to select the measurement point by the averaging
factor instead of the measurement point number.
9. Test signal current level monitor value
Displays the monitor value for the current level applied to the DUT in the present
measurement.
10. Test signal voltage level monitor value
Displays the monitor value for the voltage level applied to the DUT in the present
measurement.
11. DC resistance (Rdc) measurement value
When the DC resistance (Rdc) measurement function is ON, this measurement result is
indicated here.
A limit for making “pass or fail” judgment can be set for the DC resistance (Rdc)
measurement result, and this can be used for checking the contact condition of the
measurement electrodes and the DUT in the case of impedance measurement.
NOTE
Because the 4287A's DC resistance (Rdc) measurement function utilizes a two-terminal
method, it is unsuitable for low resistance measurement.
12. Softkey menu name
Displays the generic softkey menu name for a set of the softkeys described in “13.
Softkeys” area.
Summary of Contents for Twelfth 4287A RF LCR
Page 8: ...8...
Page 36: ...36 Chapter2 Installation Guide Preparation for Power Supply Figure 2 17 Power cable options...
Page 114: ...114 Chapter5 Setting Measurement Conditions Setting up dc resistance Rdc measurement...
Page 250: ...250 Chapter11 Setting and Using the Control and Management Facilities System Recovery...
Page 294: ...294 AppendixA Manual Changes Manual Changes...
Page 328: ...328 AppendixF Error Messages Undefined header...