136
Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Step
Test Function
Range
5520A Output
Error from nominal 1 year
U1251A
U1252A
19
Press
button to go to
A mode
5 A
5 A, 1 kHz
± 42 mA
± 37 mA
3A
3 A, 5 kHz
± 96 mA
± 96 mA
10 A
[11]
10 A, 1 kHz
± 100 mA
± 90 mA
Square Wave
Output
Use 53131A
20
Turn the rotary switch to the
position
120 Hz @ 50%
N/A
± 26 mHz
4800 Hz @ 50%
N/A
± 260 mHz
Duty Cycle
100 Hz @ 50%
N/A
± 0.398%
[12]
100 Hz @ 25%
N/A
± 0.398%
[12]
100 Hz @ 75%
N/A
± 0.398%
[12]
Use 34410A
Amplitude
4800 Hz @
99.609%
N/A
± 0.2V