150
Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Step
Test Function
Cal Range
Input
Cal Item
U1251A
U1252A
5
Press
button to go to the
mV mode
50mV
3 mV, 1 kHz
03.000 mV
03.000 mV
30 mV, 1 kHz
30.000 mV
30.000 mV
30 mV, 50 kHz
N/A
30.000 mV
500mV
30 mV, 1 kHz
030.00 mV
030.00 mV
300 mV, 1 kHz
300.00 mV
300.00 mV
300 mV, 50 kHz
N/A
300.00 mV
1000mV
30 mV, 1 kHz
0030.0 mV
0030.0 mV
1000 mV, 1 kHz
1000.0 mV
1000.0 mV
1000 mV, 50 kHz
N/A
1000.0 mV
6
Turn the rotary switch to the
W
position
Short
Dual Banana
Plug with copper
wire short
between 2
terminals
SHort
SHort
50 M
W
Input terminal
open (Remove
any test leads
and Short Plugs
from the input
terminal)
oPEn
oPEn
10 M
W
10.000 M
W
10.000 M
W
5 M
W
3 M
W
3.0000 M
W
3.0000 M
W
500 k
W
300 k
W
300.00 k
W
300.00 k
W
50 k
W
30 k
W
30.000 k
W
30.000 k
W
5 k
W
3k
W
3.0000 k
W
3.0000 k
W
500
W
300
W
300.00
W
300.00
W