152
Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
10
Press
button to go to
500
m
A
30
m
A,1 kHz
030.00
m
A
030.00
m
A
m
A mode
300
m
A,1 kHz
300.00 uA
300.00
m
A
5000
m
A
300
m
A,1 kHz
0300.0
m
A
0300.0
m
A
3000
m
A, 1 kHz
3000.0
m
A
3000.0
m
A
11
Turn the Rotary Switch to the
mA.A position
Open
Input terminal
open (Remove any
test leads and
Short Plugs from
the input terminal)
oPEn
oPEn
50 mA
30 mA
30.000 mA
30.000 mA
440 mA
300 mA
300.00 mA
300.00 mA
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution:Connect calibrator to hanheld multimeter's A and COM terminal before applying 3A and 10A
5 A
3 A
3.0000 A
3.0000 A
10 A
10 A
10.000 A
10.000 A
Move the test lead from A and COM terminal to uA.mA and COM terminal
12
Press
button to go to
mA mode
50 mA
3 mA, 1 kHz
03.000 mA
03.000 mA
30 mA, 1 kHz
30.000 mA
30.000 mA
440 mA
30 mA, 1 kHz
030.00 mA
030.00 mA
300 mA, 1 kHz
300.00 mA
300.00 mA
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution:Connect calibrator to hanheld multimeter's A and COM terminal before applying 3A and 10A
13
Press
button to go to
A mode
5 A
0.3 A, 1 kHz
0.3000 A
0.3000 A
3 A, 1 kHz
3.0000 A
3.0000 A
10 A
3 A, 1 kHz
3.0000 A
3.0000 A
10 A, 1 kHz
10.000 A
10.000 A
Summary of Contents for U1251A
Page 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Page 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Page 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Page 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...