124
Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures is listed below. If the
exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternative method would be to use the Agilent
3458A 8½ - Digit Digital Multimeter to measure less accurate
yet stable sources. The output value measured from the
source can be entered into the instrument as the target
calibration value.
Table 4
Recommended Test Equipment
Application
Recommended Equipment
Recommended Accuracy Requirements
DC Voltage
Fluke 5520A
<1/5 instrument 1 year spec
DC Current
Fluke 5520A
<1/5 instrument 1 year spec
Resistance
Fluke 5520A
<1/5 instrument 1 year spec
AC Voltage
Fluke 5520A
<1/5 instrument 1 year spec
AC Current
Fluke 5520A
<1/5 instrument 1 year spec
Frequency
Agilent 33250A
<1/5 instrument 1 year spec
Capacitance
Fluke 5520A
<1/5 instrument 1 year spec
Duty Cycle
Fluke 5520A
<1/5 instrument 1 year spec
Nanosiemens
Fluke 5520A
<1/5 instrument 1 year spec
Diode
Fluke 5520A
<1/5 instrument 1 year spec
Frequency
Counter
Agilent 33250A
<1/5 instrument 1 year spec
Temperature
Fluke 5520A
<1/5 instrument 1 year spec
Square Wave
Agilent 53131A and Agilent 34401A
<1/5 instrument 1 year spec
Short
Shorting Plug - Dual banana plug with copper
wire short between 2 terminal
<1/5 instrument 1 year spec
Summary of Contents for U1251A
Page 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Page 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Page 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Page 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...