126
Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Current Terminal Test
This test determines if the input warning of the current ter-
minal test is functioning properly.
The meter sounds an alert beep when the test lead is
inserted to A terminal but rotary switch is not set to mA.A
function. The primary display will indicate “A- Err”. It is
shown in Figure 55. The primary display will keep flashing
unless the test lead is removed from “A” terminal.
Figure 55 Input warning
N O T E
Before conducting this test, make sure the beep function is not disabled in
setup.