Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Now you can use 1234 as the security code. If want to enter
a new security code, see
“To Change the Instrument Calibra-
tion Security Code from the Front Panel”
on page 140. Make
sure you record the new security code.
Calibration Process
The following general procedure is the recommended method
to complete a full instrument calibration.
1
Read
2
Perform the verification tests to characterize the
instrument (incoming data).
3
Unsecure the instrument for calibration (see
4
Perform the adjustment procedures (see
5
Secure the instrument against calibration.
6
Note the new security code and calibration count in the
instrument's maintenance records.
N O T E
Make sure to quit the Adjustment Mode then turn off the instrument.