Performance Tests and Calibration
7
Agilent U1251A/U1252A User’s and Service Guide
145
Valid Adjustment Input Values
Adjustment can be accomplished using the following input
values below.
Table 6
Valid adjustment input values
Function
Range
Valid Amplitude Input Values
5V, 50 V, 500 V, 1000 V
0.9 to 1.1 x Full Scale
V
(for U1251A)
5 V, 50 V, 500 V, 1000 V
0.9 to 1.1 x Full Scale
(for U1252A)
5 V, 50 V ,500 V, 1000 V
0.9 to 1.1 x Full Scale
50 mV, 500 mV, 1000 mV
0.9 to 1.1 x Full Scale
500
µ
A, 5000
µ
A
0.9 to 1.1 x Full Scale
50 mA, 440 mA, 5 A, 10 A
0.9 to 1.1 x Full Scale
Ω
500
Ω
, 5k
Ω
, 50 k
Ω
,500 k
Ω
, 5M
Ω
,
50 M
Ω
0.9 to 1.1 x Full Scale
/ TEMP
10 nF, 100 nF, 1000 nF, 10
µ
F, 100
µ
F,
1000
µ
F, 10 mF / 0
°
C
Make sure to provide 0
°
C with
ambient compensation