Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Adjustment Procedure
Review the
on page 144 sections before
beginning this procedure.
1
Turn the rotary switch to “Test Function” position, shown
in the adjustment table.
2
After unsecuring the instrument, the instrument goes into
the adjustment mode.(See
“Unsecuring the Instrument for
3
The primary display will show the reference value of the
Cal Item.
4
Configure each Cal Item.
5
Use the
and
arrow keys to select the Cal
Range.
6
Apply the input signal shown in the Input column of the
table. The bar graph will display the Input reading. There
is no bar graph display for temperature adjustment.
7
Enter the actual applied input (see
8
Press
to start the adjustment. The CAL flashes in
the secondary display to indicate the calibration is in
progress.
N O T E
The instrument will be in the adjustment mode, unless Press SHIFT and
button simultaneously to exit the adjustment mode.
N O T E
Always complete tests in the same order as shown in the appropriate
table.