Specifications
DC Specifications
Table 5
DC Accuracy ± (% of r No. of Least Significant Digit)
Function
Range Resolution
Test Current/
Burden Voltage
Accuracy
U1251A
U1252A
Voltage
(1)
50.000 mV
0.001 mV
-
0.05+50
(2)
0.05+50
(2)
500.00 mV
0.01 mV
-
0.03+5
0.025+5
1000.0 mV
0.1 mV
-
5.0000 V
0.0001 V
-
50.000 V
0.001 V
-
500.00 V
0.01 V
-
0.03+5
1000.0 V
0.1 V
-
Resistance
500.00
Ω
(3)
0.01
Ω
1.04 mA
0.08+10
0.05+10
5.0000 k
Ω
(3)
0.0001 k
Ω
416
μ
A
0.08+5
0.05+5
50.000 k
Ω
0.001 k
Ω
41.2
μ
A
500.00 k
Ω
0.01 k
Ω
4.12
μ
A
5.0000 M
Ω
0.0001 M
Ω
375 nA
0.2+5
0.15+5
50.000 M
Ω
(4)
0.001 M
Ω
187 nA
1+10
1+5
500.00 M
Ω
(4)
0.01 M
Ω
187 nA
-
3+10<200M
Ω
/
8+10>200M
Ω
500.00 nS
(5)
0.01 nS
187 nA
1+20
1+10
Current
500.00
μ
A
0.01
μ
A
0.06 V (100
Ω
)
0.1+5
(6)
0.05+5
(6)
5000.0
μ
A
0.1
μ
A
0.6 V (100
Ω
)
0.1+5
(6)
0.05+5
(6)
50.000 mA
0.001 mA
0.09 V (1
Ω
)
0.2+5
(6)
0.15+5
(6)
440.00 mA
0.01 mA
0.9 V (1
Ω
)
0.2+5
(6)
0.15+5
(6)
5.0000 A
0.0001 A
0.2 V (0.01
Ω
)
0.3+10
0.3+10
10.000 A
(7)
0.001 A
0.4 V (0.01
Ω
)
0.3+10
0.3+5
Diode Test
-
0.1 mV
1.04 mA
0.05 + 5
Summary of Contents for U1251A
Page 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Page 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Page 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Page 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...