1
Calibration Procedures
Using the Front Panel for Adjustments
48
U1270 Series Service Guide
7
Use the arrow keys to enter the actual applied input
values.
8
Press
to start the adjustment.
CAL
is shown on the
display to indicate that the calibration is in progress.
9
Upon completion of each adjustment value, the display
will show
PASS
. If the adjustment fails, the instrument
will sound a long beep and the calibration error number
is shown on the display.
10
Turn the rotary switch to the next function according to
the
Test function
column shown in
Table 1- 3
. Repeat
step 3
to
step 8
for each adjustment point shown in the
adjustment table.
11
Verify the adjustments using the
“Performance Verification
Tests”
on page 27.
Exiting the adjustment mode
1
Remove all the shorting plugs and connectors from the
instrument.
2
Record the new Calibration Count.
3
Press
and
simultaneously to exit the
Adjustment Mode.
4
Cycle the instrument’s power. The instrument will then be
secured.
N O T E
You are highly recommended to complete the adjustments in the same
order as shown in
Table 1-3
.
Hz % ms
Log
N O T E
If the adjustment fails, check the input value, range, function, and entered
adjustment value before repeating the adjustment steps.
Shift
View
Esc
Auto
Trig
Hold
Summary of Contents for U1270 Series
Page 1: ...Agilent Technologies Agilent U1270 Series Handheld Digital Multimeters Service Guide...
Page 10: ...X U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 14: ...XIV U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 16: ...XVI U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...