Specifications
8
U1401A User’s and Service Guide
149
[1]
Crest factor:
≤
3
AC+DC specifications
[1]
Input impedance: 1.1 M
Ω
in parallel with <100 pF (nominal) for the range of 5 V and above, and 1 G
Ω
(nominal) for the
50/500 mV range. Crest factor:
≤
3
Table 8-4
AC current specifications
Function
Range
Resolution
Accuracy
45 Hz to 5 kHz
Burden voltage/shunt
Overload protection
AC current
[1]
(True-rms: From
5% to 100% of
range)
50 mA
1
μ
A
0.6% + 20
0.06 V (1
Ω
)
250 V, 630 mA
Quick acting fuse
500 mA
10
μ
A
0.6 V (1
Ω
)
Table 8-5
AC+DC mV/voltage specifications
Function
Range
Resolution
Accuracy
Overload protection
45 Hz to 5 kHz
5 kHz to 20 kHz
AC+DC mV/
voltage
[1]
(True-rms: From 5%
to 100% of range)
50 mV
1
μ
V
0.8% + 70
1.6% + 70
250 Vrms
500 mV
10
μ
V
0.8% + 25
1.6% + 25
5 V
0.1 mV
50 V
1 mV
250 V
10 mV
Summary of Contents for U1401A
Page 1: ...Agilent Technologies U1401A Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401A User s and Service Guide Contents ...
Page 46: ...26 U1401A User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401A User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401A User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401A User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401A User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401A User s and Service Guide 6 Maintenance ...