Specifications
8
U1401A User’s and Service Guide
151
Frequency specifications
Frequency sensitivity and trigger level for voltage measurement
For the maximum input voltage- frequency product (V- Hz)
and input impedance, refer to AC voltage measurement.
Table 8-8
Frequency specifications
Range
Resolution
Accuracy
Minimum input
frequency
Overload protection
100 Hz
0.001 Hz
0.02% + 3
1 Hz
250 Vrms
1 kHz
0.01 Hz
10 kHz
0.1 Hz
100 kHz
1 Hz
200 kHz
10 kHz
Table 8-9
Frequency sensitivity and trigger level specifications for voltage measurement
Input range
(Maximum input for specified
accuracy = 10 × range or 250 V)
Minimum sensitivity
(rms sine wave)
Trigger level for DC coupling
1 Hz to 100 kHz
>100 kHz
<20 kHz
20 kHz to 200 kHz
50 mV
15 mV
25 mV
20 mV
30 mV
500 mV
35 mV
50 mV
60 mV
80 mV
5 V
0.3 V
0.5 V
0.6 V
0.8 V
50 V
3 V
5 V
6 V
8 V
250 V
30 V
—
60 V
—
Summary of Contents for U1401A
Page 1: ...Agilent Technologies U1401A Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401A User s and Service Guide Contents ...
Page 46: ...26 U1401A User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401A User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401A User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401A User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401A User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401A User s and Service Guide 6 Maintenance ...