Specifications
8
U1401A User’s and Service Guide
153
1 ms peak hold specifications
Resistance specifications
The following resistance specifications are valid if the
maximum open voltage is less than +4.8 V. For continuity
test, the instrument will beep when the resistance is less
than 10.00
Ω.
[1]
The accuracy of 500
Ω
and 5 k
Ω
is specified after applying the relative function, which is used to offset the test lead
resistance and the thermal effect.
[2]
For the 50 M
Ω
range, the R.H. is specified for <60%.
Table 8-13
Peak hold specifications
Signal width
Accuracy for DC mV/voltage/current
Single event >1 ms
2% + 400 for all ranges
Table 8-14
Resistance specifications
Range
Resolution
Accuracy
Minimum input
current
Overload protection
500
Ω
[1]
0.01
Ω
0.15% + 8
0.45 mA
250 V rms
5 k
Ω
[1]
0.1
Ω
0.15% + 5
0.45 mA
50 k
Ω
1
Ω
45
μ
A
500 k
Ω
10
Ω
4.5
μ
A
5 M
Ω
0.1 k
Ω
450 nA
50 M
Ω
[2]
1 k
Ω
1% + 8
45 nA
Summary of Contents for U1401A
Page 1: ...Agilent Technologies U1401A Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401A User s and Service Guide Contents ...
Page 46: ...26 U1401A User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401A User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401A User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401A User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401A User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401A User s and Service Guide 6 Maintenance ...