Calibrator Output Operations
2
U1401A User’s and Service Guide
35
Defining the autoscan parameters in the memory
Press and hold
MODE
for more than one second to enter the
autoscan adjustment mode. A total of 16 steps with
individually definable time interval and amplitude are
available.
When the instrument is in the autoscan adjustment mode,
the secondary display shows the amplitude. The first two
digits of the primary display are used to indicate which step
is being adjusted. The last two digits of the primary display
are used to indicate time interval.
1
Press
MODE
to cycle through step, time interval, and
amplitude adjustments. The digit to be adjusted will flash
on the display.
•
For amplitude adjustment, press
and
to select
the digit to be adjusted, then press
and
to
adjust the value of the selected digit. The amplitude
can be set to any value within the selected output
range (±1.5 V or ±15 V for constant voltage output, ±25
mA for constant current output).
•
For time interval adjustment, press
and
to
select the digit to be adjusted, then press
and
to adjust the value of the selected digit. The time
interval can be set within the range of 0 to 99 seconds.
•
Press
for more than one second to directly reset
the time interval and amplitude of the present step to
zero.
3
4
5
6
3
4
5
6
4
Summary of Contents for U1401A
Page 1: ...Agilent Technologies U1401A Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401A User s and Service Guide Contents ...
Page 46: ...26 U1401A User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401A User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401A User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401A User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401A User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401A User s and Service Guide 6 Maintenance ...