background image

U1401A Quick Start Guide

Performing Resistance Measurement and Continuity Test

Performing Temperature Measurements

C A U T I O N

Disconnect circuit power and discharge all high-voltage capacitors before measuring 
resistance to prevent possible damage to the instrument or the device under test.

WA R N I N G

• Before measuring the temperature of a circuit or device, disconnect its power.

• The bead type thermocouple probe is suitable for measuring temperature from 

–40 °C to 204 °C in PTFE compatible environments. The probes may emit toxic 
gas above this temperature range.

1

Set the rotary switch to 

.

2

Connect the red (+) and black 
(–) test leads to the 

INPUT

 

terminals.

3

Probe the resistor (or shunt) 
leads and read the display.

4

To perform continuity test, 
press 

 to toggle the 

audible continuity function 

on

 

or 

off

.

For continuity test, the instrument will 
beep if the resistance is less than the 
values indicated below:

Measurement range

Resistance 

threshold

500.00 

Ω

10 

Ω

5.0000 k

Ω

100 

Ω

50.000 k

Ω

1 k

Ω

500.00 k

Ω

10 k

Ω

5.0000 M

Ω

100 k

Ω

50.000 M

Ω

1 M

Ω

1

Set the slide switch to the 

M

 

position to disable the output.

2

Set the rotary switch to 

mV

.

3

Press and hold 

 for more 

than 1 second.

4

Plug the thermocouple probe 
(with adapter) into the 

INPUT

 

terminals.

5

Touch the surface to be 
measured with the 
thermocouple probe and read 
the display.

Best practices:

Do not bend the thermocouple 
leads at sharp angles. 
Repeated bending may break 
the leads.

Do not immerse the 
thermocouple probe in any 
liquid.

Clean the surface to be 
measured and make sure that 
the probe is securely touching 
the surface.

U1401A QSG.fm  Page 3  Tuesday, May 15, 2012  11:10 AM

Summary of Contents for U1401A

Page 1: ...ely and the Agilent GUI Data Logger Software downloadable from www agilent com find hhTechLib WARNING Ensure that the terminal connections are correct for that particular measurement selection before starting any measurement To avoid damage to the device do not exceed the rated input limit CAUTION Degradation of some product specifications can occur in the presence of ambient electromagnetic EM fi...

Page 2: ...nts and read the display 4 Press to display dual measurements Measuring DC voltage 1 Set the rotary switch to V or mV Press to select 2 Connect the red and black test leads to the INPUT terminals 3 Probe the test points and read the display 4 Press to display dual measurements Measuring AC current 1 Set the rotary switch to mA Press to select 2 Connect the red and black test leads to the INPUT ter...

Page 3: ...To perform continuity test press to toggle the audible continuity function on or off For continuity test the instrument will beep if the resistance is less than the values indicated below Measurement range Resistance threshold 500 00 Ω 10 Ω 5 0000 kΩ 100 Ω 50 000 kΩ 1 kΩ 500 00 kΩ 10 kΩ 5 0000 MΩ 100 kΩ 50 000 MΩ 1 MΩ 1 Set the slide switch to the M position to disable the output 2 Set the rotary ...

Page 4: ...o any one of the positions 2 Press The annunciator appears 3 Press to cycle through 1 5 V 15 V 1 5 V 15 V 1 5 V and 15 V output modes 4 Select 1 5 V or 15 V 5 Press to output the signal Constant current output 1 Set the rotary switch to any one of the positions 2 Press The annunciator appears 3 Press to cycle through 25 mA 25 mA and 25 mA output modes 4 Select 25 mA 5 Press to output the signal 1 ...

Page 5: ...urns on autorange Press and hold for 1 s Turns on dual display Press Turns on the frequency measurement function Press Enables the shifted functions of the keypad Press Toggles the backlight on or off Press and hold for 1 s Selects output mode and range for voltage and current output With rotary switch at or and with mode enabled press Selects parameter adjustments for square wave output With rota...

Page 6: ...rge cycles before achieving maximum capacity To discharge simply run the multimeter using the rechargeable battery s power until it shuts down or until the low battery warning appears In some instances the multimeter may indicate that the charging is complete after ten minutes or so when charging a new rechargeable battery This is a normal phenomenon with rechargeable batteries Remove the recharge...

Page 7: ...formed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met WARNING A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond...

Page 8: ...Agilent Technologies Second Edition May 4 2012 Agilent Technologies Inc 2008 2012 U1401 90011 Printed in Malaysia U1401A QSG fm Page 8 Tuesday May 15 2012 11 10 AM ...

Reviews: