background image

U1731A/U1732A Quick Start Guide

6

Figure 1-1

Inductance measurement

Figure 1-1

Capacitance measurement

Figure 1-1

Resistance measurement

Summary of Contents for U1731A

Page 1: ...Agilent Technologies Agilent U1731A U1732A Dual Display Handheld LCR Meter Quick Start Guide...

Page 2: ...until the indicated conditions are fully understood and met WARNING WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not cor rectly performed...

Page 3: ...fuse NOTE Your multimeter is capable of remote data logging To use this feature you will need an IR USB cable U5481A purchased separately and the Agilent GUI Data Logger Software downloadable from www...

Page 4: ...e to set 1 5 10 and 20 for sorting capacitance 19 Unit for Beeper Frequency as setup mode 7 AUTO range 20 Parallel mode indicator 8 Inductance Capacitance or Resistance L C or R function indicator 21...

Page 5: ...t To toggle parallel and series mode 12 GUARD Guard terminal socket 4 TOL Tolerance mode 5 FREQ To select test frequency 6 REL CAL Relative mode CAL function mode 7 Manual AUTO Manual range Auto range...

Page 6: ...Press the D Q key to select Q or D factor for secondary display as required 6 Ensure that your hands are not touching the clips before taking the readings on the display 7 The LCR measurements setup a...

Page 7: ...U1731A U1732A Quick Start Guide 6 Figure 1 1 Inductance measurement Figure 1 1 Capacitance measurement Figure 1 1 Resistance measurement...

Page 8: ...Hz 100 Hz 120 Hz 120 Hz 1 kHz 1010 Hz 10 kHz 9 6 kHz Measuring rate 1 time s nominal Response time 1 s manual range Auto power off 5 mins without operations Low battery indicator 6 8 V Operating tempe...

Page 9: ...roduct Reference CD ROM Agilent U1731A U1732A Quick Start Guide Alligator clip leads 9 V alkaline battery Certificate of Calibration Optional accessories IR to USB cable ordering option U5481A FG Powe...

Page 10: ...9 999 MW 2 0 8 2 0 8 This specification is based on the battery operation After open cal 2000 kW 1999 9 kW 0 5 5 0 5 5 After open cal 200 kW 199 99 kW 0 5 3 0 5 3 20 kW 19 999 kW 0 5 3 0 5 3 2000 W 19...

Page 11: ...00 Cx 5 DF 0 1 After open cal Range Maximum Display Accuracy Specified Note Capacitance DF 1 mF 1 999 mF This reading can be extended up to 1999 MAX display with accuracy that is not specified 3 0 5 D...

Page 12: ...10000 5 10 0 100 Lx 5 After short cal Range Maximum Display Accuracy Specified Note Inductance DF 100 H 99 99 H 1 0 Lx 10000 5 2 0 100 Lx 5 After open cal 20 H 19 999 H 0 7 Lx 10000 5 1 2 100 Lx 5 200...

Page 13: ...0 5 3 0 5 3 20 kW 19 999 kW 0 5 3 0 5 3 2000 W 1999 9 W 0 5 3 0 5 3 200 W 199 99 W 0 8 5 0 8 5 After short cal 20 W 19 999 W 1 2 40 1 2 40 After short cal Range Maximum Display Accuracy Specified Not...

Page 14: ...0 7 3 DF 0 5 0 7 100 Cx 5 DF 0 5 2000 nF 1999 9 nF 0 7 3 DF 0 5 0 7 100 Cx 5 DF 0 5 200 nF 199 99 nF 0 7 5 DF 0 5 0 7 100 Cx 5 DF 0 5 After open cal 20 nF 19 999 nF 1 0 5 DF 0 1 2 0 100 Cx 5 DF 0 1 A...

Page 15: ...1 5 0 100 Cx 8 DF 0 1 After open cal NOTE 1 Q value is the reciprocal of DF 2 This specification is based on the measurement performed at the test socket 3 DUT and test leads need to be properly shie...

Page 16: ...00 5 3 0 100 Lx 5 After short cal 2000 H 1999 9 H 2 0 Lx 10000 5 10 0 100 Lx 5 After short cal Range Maximum Display Accuracy Specified Note Inductance DF 1000 mH 999 9 mH 2 0 Lx 10000 8 2 0 100 Lx 10...

Page 17: ...U1731A U1732A Quick Start Guide 16...