11
SENSe Subsystem
112
U2300A Series USB DAQ
[SENSe:]COUNter:GATE:CONTrol
Syntax
[SENSe:]COUNter:GATE:CONTrol <mode>, (@<ch_list>)
This command is used set the gate status (enable/disable) of the general
purpose digital counter
.
[SENSe:]COUNter:GATE:CONTrol? (@<ch_list>)
The query command returns a string value representing the gate status on
each specified counter channel. Multiple responses are separated by
comma.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Query Returned Format
<mode> ::= {DIS|ENAB}
See Also
[SENSe:]COUNter:TOTalize:CLEar
Item
Type
Range of Values
Default Value
<mode>
String
DIS: Stop counter operation
ENAB: Start counter operation
DIS
<ch_list>
Numeric
301 through 302
N/A
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...