15
WAVeform Subsystem
148
U2300A Series USB DAQ Programming Guide
WAVeform:DATA?
Syntax
WAVeform:DATA?
This command is used to returns the raw data of input channels that are
enabled by
ROUTe:SCAN
command. For example,
#800000200<byte
1><byte 2>…<byte 200>
, indicates that there is a total of 200 bytes of
returned data. Each acquisition is made up of two bytes. The first byte is
the LSB and the second one is the MSB. The sample shows that there is a
total of 100 acquisition points.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Remarks
•
Maximum 8 Msa in
DIGitize
or 4 Msa in
RUN
.
•
The length of header is fixed at eight characters.
Example
*RST; *CLS
ROUTe:SCAN (@101)
//Monitors only channel 101
ACQuire:SRATe 30000
//Set sampling rate of 30 KHz
WAVeform:POINts 3000
//Set to read/fetch 3000 points at one time
RUN
WAVeform:STATus?
//Expect a string returned data
WAVeform:DATA?
//Loop until the next STOP command is issued
STOP
//Acquisition stops
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...