MEASure Subsystem
7
U2300A Series USB DAQ Programming Guide
55
MEASure:COUNter:DATA?
Syntax
MEASure:COUNter:DATA? (@<ch_list>)
The query command returns a float value based on the
SENSe:COUNter
command. This value is the counter value for specific counting mode on
specified counter channel. Multiple responses are separated by commas.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Remarks
This query requires a channel number to be stated.
Query Returned Format
<value>
Example
SENSe:COUNter:FUNCtion TOTalize, (@301)
SENSe:COUNter:FUNCtion FREQuency, (@301)
MEASure:COUNter:DATA? (@301)
See Also
Item
Type
Range of Values
Default Value
<ch_list>
Numeric
301 through 302
N/A
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...