ROUTe Subsystem
10
U2300A Series USB DAQ Programming Guide
87
Remarks
•
For U2300A Series DAQ device that supports multiplexing only.
•
The maximum channel list is 100 items.
•
Repeated channel assignment is allowed, for instance 101, 102, 103,
101…
•
This command is capable of changing the scan order without changing
the properties on RANGE, POLARITY and SIGNAL TYPE.
•
Consecutive channels can be specified using colon or comma. For
example 101, 102, 103 or 101:103.
Query Returned Format
(@<ch_list>)