ACQuire Subsystem
2
U2300A Series USB DAQ Programming Guide
13
ACQuire:POINts
Syntax
ACQuire:POINts <value>
This command is used to set the count of a single- shot analog input.
ACQuire:POINts?
The query command returns a numeric value, representing the acquisition
count for single- shot analog input. Note that the single- shot acquisition is
initiated by the
DIGitize
command.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Remarks
<value> * <number of entries>
is equivalent to eight mega points.
Query Returned Format
<value>
See Also
Item
Type
Range of Values
Default Value
<value>
Numeric
Maximum of 8 Msa
500