8
OUTPut Subsystem
66
U2300A Series USB DAQ Programming Guide
OUTPut:WAVeform:SRATe
Syntax
OUTPut:WAVeform:SRATe <value>
This command is used to set the updated rate of analog outputs.
OUTPut:WAVeform:SRATe?
The query command returns an integer value representing the updated
rate of analog outputs measured in Hz.
Supported Model
U2351A, U2353A, U2355A, U2356A, U2331A
Parameters
Remarks
•
The range bound by 0
≤
<value>
≤
10
6
. Must be
≥
48x10
6
/ 0xffffff.
•
Default value zero implies automatic sampling configuration for
SIN
,
SQU
,
SAWT
,
TRI
, and
NOIS
waveform generation.
•
Maximum of 4096 points for one waveform.
Query Returned Format
<value> [Hz]
Name
Type
Range of Values
Default Value
<value>
Float
(48/16777215) MHz to 1 MHz
0
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...