Specifications
8
U8903A User’s Guide
133
Sweep capability
FREQUENCY SWEEP (SINE AND DUAL SINE WAVEFORMS)
Start point
5 Hz to 80 kHz
Stop point
5 Hz to 80 kHz
Amplitude spot value
(Balanced)
0 V to 16 Vrms
Amplitude spot value
(Unbalanced/Common)
0 V to 8 Vrms
FREQUENCY SWEEP (SQUARE WAVEFORM)
Start point
5 Hz to 30 kHz
Stop point
5 Hz to 30 kHz
Amplitude spot value
(Balanced)
0 V to 22.6 Vrms
Amplitude spot value
(Unbalanced/Common)
0 V to 11.3 Vrms
FREQUENCY SWEEP (SMPTE IMD 1:1/4:1/10:1 WAVEFORM)
Start point (Upper frequency)
2 kHz to 60 kHz
Start point (Lower frequency)
40 Hz to 500 Hz
Stop point (Upper frequency)
2 kHz to 60 kHz
Stop point (Lower frequency)
40 Hz to 500 Hz
Amplitude spot value
(Balanced)
0 V to 16 Vrms
Amplitude spot value
(Unbalanced/Common)
0 V to 8 Vrms
FREQUENCY SWEEP (DFD IEC 60118/IEC 60268 WAVEFORM)
Start point (Upper frequency)
3 kHz to 80 kHz
Start point (Center frequency)
3 kHz to 79 kHz
Stop point (Upper frequency)
3 kHz to 80 kHz
Stop point (Center frequency)
3 kHz to 79 kHz
Amplitude spot value
(Balanced)
0 V to 16 Vrms
Amplitude spot value
(Unbalanced/Common)
0 V to 8 Vrms
Difference frequency spot value
80 Hz to 2 kHz
Summary of Contents for u8903a
Page 1: ...Agilent Technologies Agilent U8903A Audio Analyzer User s Guide...
Page 12: ...XII U8903A User s Guide...
Page 20: ...XX U8903A User s Guide...
Page 22: ...XXII U8903A User s Guide...
Page 38: ...16 U8903A User s Guide 1 Getting Started...
Page 74: ...52 U8903A User s Guide 2 Operation and Features...
Page 90: ...68 U8903A User s Guide 3 Instrument Configuration...