134
U8903A User’s Guide
8
Specifications
VOLTAGE SWEEP
Start point (Balanced)
0 V to 16 Vrms
[1]
Start point
(Unbalanced/Common)
0 V to 8 Vrms
Stop point (Balanced)
0 V to 16 Vrms
Stop point
(Unbalanced/Common)
0 V to 8 Vrms
Frequency spot value (Sine and
Dual sine waveforms)
5 Hz to 80 kHz
Frequency spot value (Square
waveform)
5 Hz to 30 kHz
Frequency spot value (SMPTE
IMD 1:1/4:1/10:1 waveform)
40 Hz to 500 Hz (lower frequency), 2 kHz to 60 kHz (upper frequency)
Frequency spot value (DFD IEC
60118/IEC 60268 waveform)
3 kHz to 80 kHz (upper frequency), 3 kHz to 79 kHz (center frequency), 80 Hz to 2 kHz
(difference frequency)
DC SWEEP
Start point (Balanced)
–22.6 V to 22.6 V
Start point
(Unbalanced/Common)
–11.3 V to 11.3 V
Stop point (Balanced)
–22.6 V to 22.6 V
Stop point
(Unbalanced/Common)
–11.3 V to 11.3 V
PHASE SWEEP
Start point
–180
to 179.99
Stop point
–180
to 179.99
[1] This range is applicable for sine wave only.