5 Scale Calibration
The scale is configured from the factory to match the specified settings for each unit,
as defined by the product specifications and sales brochure. Modification of the
settings can be accomplished by altering user configuration settings P7-P10.
5.1 Setup Controls
5.2 User Configuration Settings
CAUTION!
Calibration and/or configuration of calibration settings of your scale should be
accomplished by a trained service technician using certified weights to ensure
proper operation and accuracy. Calibration is not covered under warranty.
Moves flashing digit
Change flashed digits
Saves data and move down to next
parameter setting
Exits setup mode
Hold
Units
Tare
On/Off
Zero
kg
kg
Config Menu Avail. Settings
Definition
Detailed Setting
P7
00-31
Displayed Resolution
(00) = 500
(01)= 600
(02)= 750
(03)= 800
(04) =1000
(05)=1200
(06) =1500
(07) = 2000
(08) = 2400
(09) = 2500
(10)= 3000
(11) = 3500
(12)= 4000
(13) = 5000
(14) = 6000
(15) = 7000
(16) = 7500
(17) = 8000
(18) = 10,000
(19) = 12,000
(20) = 15,000
(21-31) = N/A
P8
0,1,2
Division Increment
0 = 1; 1 = 2; 2 = 5
P9
0-5
Decimal Position
0 = 123456; 1 = 12345.6; 2 = 1234.56;
3 = 123.456; 4 = 12.3456; 5 = 1.23456
P10
0,1
Calibration weight
0 = kg; 1 = lb
Summary of Contents for Brecknell SBI-100
Page 1: ...SBI 100 User Instructions AWT 35 501281 Issue AA ...
Page 4: ...2 SBI 100 User Instructions Manual Chapter 7 Error Codes 29 ...
Page 8: ...6 SBI 100 User Instructions Manual 1 4 Declaration of Compliance ...
Page 27: ...SBI 100 User Instructions Manual 25 7 8 9 Tare Tare ZERO 0 ...
Page 32: ...30 SBI 100 User Instructions Manual ...
Page 33: ......