10
© Agrowtek Inc. | www.agrowtek.com
1. Select
CALIBRATE
from the conductivity
calibration menu.
3. When the reading is
stable, press DONE to
complete the calibration.
2. Follow the instruc-
tions then press
NEXT
to continue.
DWN ENT
UP
BACK
CALIBRATE
ADVANCED
CALIBRATION
GPH MNU
H/L
72.0
6.10
1550
600
°F
pH
uS
mV
DWN ENT
UP
EXIT
ALARMS
CALIBRATION
SETUP
PUMPS
MENU
Conductivity Calibration
Conductivity calibration is a single point process requiring 1413 uS calibration solution. The temperature
probe must be inserted into the calibration solution at the same time as the conductivity probe.
NEXT
EXIT
CONDUCTIVITY
LAST CALIBRATION
10/19/2017
PUT TEMPERATURE
AND EC PROBE IN
1413uS
SOLUTION.
1. Select
CALIBRATE
from the conductivity
calibration menu.
3. When the reading is
stable, press DONE to
complete the calibration.
2. Follow the instruc-
tions then press
NEXT
to continue.
DWN ENT
UP
BACK
CALIBRATE
ADVANCED
CALIBRATION
GPH MNU
H/L
72.0
6.10
1550
600
°F
pH
uS
mV
DWN ENT
UP
EXIT
ALARMS
CALIBRATION
SETUP
PUMPS
MENU
O.R.P. Calibration
ORP (optional) calibration is a single point process requiring 470 mV calibration solution.
NEXT
EXIT
ORP
LAST CALIBRATION
10/19/2017
PUT ORP PROBE IN
470mV
SOLUTION.
DNE
EXIT
CONDUCTIVITY
WAIT FOR READING
TO STABILIZE THEN
PRESS DONE.
1440 uS
DWN ENT
UP
BACK
TIME/DATE
UNITS
LOG INTERVAL
SETUP
NEXT
DNE
EXIT
CONDUCTIVITY
WAIT FOR READING
TO STABILIZE THEN
PRESS DONE.
1440 uS
DWN ENT
UP
BACK
TEMPERATURE
PH
COND
CALIBRATION
DO
DWN ENT
UP
BACK
ORP
CLEAR ALL
CALIBRATION