background image

13

© Agrowtek Inc. | www.agrowtek.com

Advanced Calibration

Sensors values may be manually calibrated to alternate standards using the 

advance calibration features. 

OFFSET calibration applies a linear offset adjustment to the value. 

SPAN calibration applies an adjustment to the slope of the sensor value.

Note:  pH and conductivty are temperature compensated (ATC) sensors. For 

accurate calibration, the temperature probe must be in the pH and conduc-

tivity calibration standards. Allow all readings to stabilize before performing 

an offset or span calibration operation.

PH

CAL 7 sets the pH 7 calibration (offset.)  Cal 7 automatically clears span calibration point prior performing pH 

7 calibration.

SPAN calibration is performed typically at pH 4.01 or 10.0 

(Always perform pH 7 calibration first.)

CONDUCTIVITY

SET CAL 0 calibrates a dry EC probe if required. (not recommended)

SPAN calibration is recommended at 1413uS. If using ppm standards, ensure the display is in the correct 

units.

DiSSOLVED OXYGEN (D.O.)

OFFSET calibration is recommended in zero oxygen solution.

SPAN calibration is recommended at a known DO level or in air.

OXIDATION REDUCTION POTENTIAL (O.R.P.)

OFFSET is recommended at 270mV

SPAN is recommended at 470mV

DWN ENT

UP

BACK
CALIBRATE
ADVANCED

CALIBRATION

4-20mA analog outputs may also be calibrated with a positive or negative 

offset to compensate for variation in DAC’s/ADC’s. The sensors’ current output 

may be incrementally increased or decreased in steps of 0.005mA over a range 

of +/-2mA.

1 Offset bit = 0.005mA, Range = +/-400 bits (+/-2mA)

This calibration procedure is optional and only for use with custom PLC appli-

cations.

1. Observe the PLC’s input or data readings.

2. Increase or decrease the offset value to incremenetally adjust the current output until the values match.

Analog Output Calibration

DWN ENT

UP

BACK
+1
-1

ANALOG

-5

Summary of Contents for SXH

Page 1: ...ture Calibration 9 pH Calibration 9 Conductivity Calibration 10 O R P Calibration 10 D O Calibration 11 D O Saturation Chart 12 Advanced Calibration 13 Analog Output Calibration 13 Setup Menu 14 Time...

Page 2: ...ROWtEK U S A MADE IN Arowtek Inc www agrowtek com GrowNET Link Port MODBUS TM Digital Water Sensor Logger Controller TEMPERATURE CONDUCTIVITY pH O R P D O GrowControlTM pH Probe o r p d o Probe Temp P...

Page 3: ...pH 4 buffer Take out electrode by loosening plastic top on bottle counter clockwise and pulling electrode out Slide cap and O ring off electrode and save BLACK O RING INSIDE TOP Black O ring inside A...

Page 4: ...ure out the hole locations per the dimensions or mark the holes using the transmitter as a template 2 Drill holes and install anchors if required not included Keep the transmitter away from dust durin...

Page 5: ...amage to the equipment 12Vdc 1A Power Supply Technology to Help You Grow AGROWtEK U S A MADE IN Arowtek Inc www agrowtek com GrowNET Link Port MODBUS TM Digital Water Sensor Logger Controller TEMPERAT...

Page 6: ...om the attached sensors Each button is labeled at the bottom of the display to describe it s function on the current screen or menu RST EXIT 72 0 4 85 657 480 F pH mV 68 4 4 21 551 404 HIGH LOW uS Tec...

Page 7: ...The out of range value will be displayed in red to indicate the cause for the alarm Additionally alarm limits are plotted on the graphs to indicate values are within the desired range Alarms Menu DWN...

Page 8: ...MENU DWN ENT UP EXIT ALARMS CALIBRATION SETUP MENU DWN ENT UP BACK TEMPERATURE PH COND CALIBRATION CLEAR ALL DWN ENT UP BACK LOW 0 0 HIGH 0 0 ALARM OFF ALARM 1 Select ALARM OFF DWN ENT UP BACK SET ON...

Page 9: ...N PRESS NEXT 7 12 pH NEXT EXIT PH STEP 2 PUT TEMPERATURE AND PH PROBE IN PH4 01 SOLUTION 7 00 pH DNE EXIT PH WAIT FOR READING TO STABILIZE THEN PRESS DONE 4 14 pH pH 7 Calibration Solution pH 4 01 Cal...

Page 10: ...TURE AND EC PROBE IN 1413uS SOLUTION 1 Select CALIBRATE from the conductivity calibration menu 3 When the reading is stable press DONE to complete the calibration 2 Follow the instruc tions then press...

Page 11: ...ion and either air or saturated solution NEXT EXIT DO LAST CALIBRATION 10 19 2017 PUT DO PROBE IN ZERO OXYGEN SOLUTION NXT EXIT DO WAIT FOR READING TO STABILIZE THEN PRESS NEXT 1 2 mg L DWN ENT UP BAC...

Page 12: ...10 3 10 1 10 0 9 9 9 8 9 6 9 5 9 4 9 3 9 2 9 1 8 9 8 8 8 7 8 6 8 5 14 10 3 10 2 10 0 9 9 9 8 9 7 9 6 9 4 9 3 9 2 9 1 9 0 8 9 8 8 8 7 8 5 8 4 8 3 15 10 1 9 9 9 8 9 7 9 6 9 5 9 3 9 2 9 1 9 0 8 9 8 8 8 7...

Page 13: ...brates a dry EC probe if required not recommended SPAN calibration is recommended at 1413uS If using ppm standards ensure the display is in the correct units DiSSOLVED OXYGEN D O OFFSET calibration is...

Page 14: ...ERVAL SETUP NEXT DWN ENT UP BACK COMM DEVICE ADD MFG INFO SETUP BACKLIGHT DWN ENT UP BACK TIME DATE UNITS LOG INTERVAL SETUP NEXT DWN ENT UP BACK TIME DATE TIME DATE EXT NXT DATE 10 20 2017 EXT NXT TI...

Page 15: ...may be displayed in F or C Note Check alarm settings when converting temperature units DWN ENT UP BACK C F UNIT SELECT 68 0 F Configure conductivity units 1 Select EC from the units menu 2 Select the...

Page 16: ...Device Address GPH MNU 0 10 50 0 F pH uS mV DWN ENT UP EXIT ALARMS CALIBRATION SETUP PUMPS MENU DWN ENT UP BACK TIME DATE UNITS LOG INTERVAL SETUP NEXT DWN ENT UP BACK COMM DEVICE ADD MFG INFO SETUP B...

Page 17: ...wtek Inc www agrowtek com WiFi AgrowNET Link Port TM WiFi Reset DigiHydro Water Monitor Controller TEMPERATURE CONDUCTIVITY pH O R P D O if equipped GrowControlTM POWER 15 24Vdc pH Probe o r p d o Pro...

Page 18: ...X8 GrowNET Hubs allow multiple GrowNET sensors relays and dosing pumps to be connected to a single LX1 or LX2 interface Indi vidually buffered full duplex ports for signal integrity Hubs can be daisy...

Page 19: ...value 700 To set the conductivity zero calibration write the value 0 Span Calibration Span or slope calibration corrects the slop of the sensor reading at a second point away from the zero cali bratio...

Page 20: ...libration Input Signed Int 16 bit W 41103 Sensor 4 Offset Cal Calibration Input Signed Int 16 bit W 41104 Sensor 1 Span Cal Calibration Input Signed Int 16 bit W 41201 Sensor 2 Span Cal Calibration In...

Page 21: ...treatments can be tried They are presented in the order of the severity of attack on the pH glass and may not improve and in some cases actually further deteriorate electrode performance DANGER Use pr...

Page 22: ...rranty Period In no event or circumstance shall Agrowtek Inc be liable to any third party or the claimant for damages in excess of the price paid for the product or for any loss of use inconvenience c...

Reviews: