15
© Agrowtek Inc. | www.agrowtek.com
Adjust the interval for recording data points in the
on-board memort. Acceptable values are from 1 -
65535 seconds.
21,600 data points can be stored for each sensor
value. The most recent 120 data points are shown
on the graphical history.
The entire data history may be downloaded from
the sensor to a .csv file with the LX1 USB AgrowLINK
and free software.
Note: 60 second intervals = 15 days of data storage.
Logging Interval
Configure temperature units:
1. Select
TEMPERATURE
from the units menu.
2. Select the desired
units and press
ENT
.
DWN ENT
UP
BACK
TEMPERATURE
EC
UNIT SELECT
Temperature may be displayed in °F or °C.
Note: Check alarm settings when converting
temperature units.
DWN ENT
UP
BACK
°C
°F
UNIT SELECT
68.0°F
Configure conductivity units:
1. Select
EC
from the
units menu.
2. Select the desired
units and press
ENT
.
Conductivity may be displayed in default units of
microSiemens (uS) or total dissolved solids in parts
per million (ppm.)
The TDS conversion factor used by this meter is 500.
TDSppm = uS x 0.5
DWN ENT
UP
BACK
TEMPERATURE
EC
UNIT SELECT
DWN ENT
UP
BACK
US
TDS
UNIT SELECT
654 ppm
GPH MNU
H/L
72.0
6.10
1550
600
°F
pH
uS
mV
DWN ENT
UP
EXIT
ALARMS
CALIBRATION
SETUP
PUMPS
MENU
DWN ENT
UP
BACK
TIME/DATE
UNITS
LOG INTERVAL
SETUP
NEXT
1. Select
LOG INTERVAL
from the setup menu.
2. Adjust the value
then select
BACK
.
DWN ENT
UP
BACK
TIME/DATE
UNITS
LOG INTERVAL
SETUP
NEXT
DWN ENT
UP
BACK
+1
-1
LOGGIN INTERVAL
60 SEC