pH & ORP Probe Preparation
Probe Shipping & Storage
pH & ORP probes are shipped in a plastic bottle containing a
solution of pH 4 buffer and potassium chloride. The electrode
should remain in the bottle until it is used.
If the electrode is used infrequently, the bottle and its solution
should be saved and the electrode stored in it. If the solution in
the soaker bottle is missing, fill the bottle with pH 4 buffer.
Take out electrode by loosening plastic top on bottle counter-
clockwise and pulling electrode out. Slide cap and O-ring off
electrode and save.
BLACK
O-RING
INSIDE TOP
Black
O-ring
inside
Air Bubble
During shipment the air bubble in the
electrode’s stem may move into the bulb
area.
If bubbles are seen in the bulb area, hold
the electrode by its top cap and shake
while pointed downward.
E
L
B B
U
B
R
I
A
AIR BUBBLE
Installation/Operation Angle
pH & ORP sensor probes must be installed in an above-horizontal position
with the probe tip facing downward to prevent the air bubble from entering
the the bulb area.
10° min.