pH calibration is a two-point process requiring both pH 7 and pH 4.01 calibration solutions. The temperature 

probe must be inserted into the calibration solution at the same time as the pH probe.

GPH MNU

H/L

72.0

6.10

1550

600

°F

pH

uS

mV

DWN ENT

UP

EXIT
ALARMS
CALIBRATION
SETUP
PUMPS

MENU

pH Calibration

2. Follow the instruc-

tions then press 

NEXT.

4. Clean the probes with DI/RO 

and change calibration solution.

3. When the reading 

is stable, press 

NEXT.

5.  When the reading is stable, press 

DONE 

to calibrate pH 4.01 and finish calibration.

NEXT

EXIT

PH

LAST CALIBRATION

10/19/2017

STEP 1:

PUT TEMPERATURE

AND PH PROBE IN

PH7

 SOLUTION.

NEXT

EXIT

PH

WAIT FOR READING

TO STABILIZE THEN

PRESS NEXT.

7.12  pH

NEXT

EXIT

PH

STEP 2:

PUT TEMPERATURE

AND PH PROBE IN

PH4.01

 SOLUTION.

7.00  pH

DNE

EXIT

PH

WAIT FOR READING

TO STABILIZE THEN

PRESS DONE.

4.14  pH

pH 7 Calibration 

Solution

pH 4.01 Calibra-

tion Solution

1. Select 

CALIBRATE 

from calibration menu.

DWN ENT

UP

BACK
CALIBRATE
ADVANCED

CALIBRATION

DWN ENT

UP

BACK
TEMPERATURE
PH
COND

CALIBRATION

DO

1. Select 

CALIBRATE 

from calibration menu.

3. Adjust to the desired 

value. Hold 

ENT

 to jog 

the value by 10x.

2. Press 

NEXT 

to con-

tinue.

4. Confirm the new 

reading or press 

NO

.

DWN ENT

UP

BACK
+0.1
-0.1

OFFSET

72.2°F

NEXT

EXIT

TEMPERATURE

LAST CALIBRATION

10/19/2017

PRESS NEXT TO

ADJUST

TEMPERATURE

READING.

NO

YES

CONFIRM?

72.2  °F

68.1  °F

OLD

NEW

DWN ENT

UP

BACK
CALIBRATE
ADVANCED

CALIBRATION

GPH MNU

H/L

72.0

6.10

1550

600

°F

pH

uS

mV

DWN ENT

UP

EXIT
ALARMS
CALIBRATION
SETUP
PUMPS

MENU

Temperature Calibration

DWN ENT

UP

BACK
TEMPERATURE
PH
COND

CALIBRATION

DO

Summary of Contents for SXH

Page 1: ...ture Calibration 9 pH Calibration 9 Conductivity Calibration 10 O R P Calibration 10 D O Calibration 11 D O Saturation Chart 12 Advanced Calibration 13 Analog Output Calibration 13 Setup Menu 14 Time...

Page 2: ...ROWtEK U S A MADE IN Arowtek Inc www agrowtek com GrowNET Link Port MODBUS TM Digital Water Sensor Logger Controller TEMPERATURE CONDUCTIVITY pH O R P D O GrowControlTM pH Probe o r p d o Probe Temp P...

Page 3: ...pH 4 buffer Take out electrode by loosening plastic top on bottle counter clockwise and pulling electrode out Slide cap and O ring off electrode and save BLACK O RING INSIDE TOP Black O ring inside A...

Page 4: ...ure out the hole locations per the dimensions or mark the holes using the transmitter as a template 2 Drill holes and install anchors if required not included Keep the transmitter away from dust durin...

Page 5: ...amage to the equipment 12Vdc 1A Power Supply Technology to Help You Grow AGROWtEK U S A MADE IN Arowtek Inc www agrowtek com GrowNET Link Port MODBUS TM Digital Water Sensor Logger Controller TEMPERAT...

Page 6: ...om the attached sensors Each button is labeled at the bottom of the display to describe it s function on the current screen or menu RST EXIT 72 0 4 85 657 480 F pH mV 68 4 4 21 551 404 HIGH LOW uS Tec...

Page 7: ...The out of range value will be displayed in red to indicate the cause for the alarm Additionally alarm limits are plotted on the graphs to indicate values are within the desired range Alarms Menu DWN...

Page 8: ...MENU DWN ENT UP EXIT ALARMS CALIBRATION SETUP MENU DWN ENT UP BACK TEMPERATURE PH COND CALIBRATION CLEAR ALL DWN ENT UP BACK LOW 0 0 HIGH 0 0 ALARM OFF ALARM 1 Select ALARM OFF DWN ENT UP BACK SET ON...

Page 9: ...N PRESS NEXT 7 12 pH NEXT EXIT PH STEP 2 PUT TEMPERATURE AND PH PROBE IN PH4 01 SOLUTION 7 00 pH DNE EXIT PH WAIT FOR READING TO STABILIZE THEN PRESS DONE 4 14 pH pH 7 Calibration Solution pH 4 01 Cal...

Page 10: ...TURE AND EC PROBE IN 1413uS SOLUTION 1 Select CALIBRATE from the conductivity calibration menu 3 When the reading is stable press DONE to complete the calibration 2 Follow the instruc tions then press...

Page 11: ...ion and either air or saturated solution NEXT EXIT DO LAST CALIBRATION 10 19 2017 PUT DO PROBE IN ZERO OXYGEN SOLUTION NXT EXIT DO WAIT FOR READING TO STABILIZE THEN PRESS NEXT 1 2 mg L DWN ENT UP BAC...

Page 12: ...10 3 10 1 10 0 9 9 9 8 9 6 9 5 9 4 9 3 9 2 9 1 8 9 8 8 8 7 8 6 8 5 14 10 3 10 2 10 0 9 9 9 8 9 7 9 6 9 4 9 3 9 2 9 1 9 0 8 9 8 8 8 7 8 5 8 4 8 3 15 10 1 9 9 9 8 9 7 9 6 9 5 9 3 9 2 9 1 9 0 8 9 8 8 8 7...

Page 13: ...brates a dry EC probe if required not recommended SPAN calibration is recommended at 1413uS If using ppm standards ensure the display is in the correct units DiSSOLVED OXYGEN D O OFFSET calibration is...

Page 14: ...ERVAL SETUP NEXT DWN ENT UP BACK COMM DEVICE ADD MFG INFO SETUP BACKLIGHT DWN ENT UP BACK TIME DATE UNITS LOG INTERVAL SETUP NEXT DWN ENT UP BACK TIME DATE TIME DATE EXT NXT DATE 10 20 2017 EXT NXT TI...

Page 15: ...may be displayed in F or C Note Check alarm settings when converting temperature units DWN ENT UP BACK C F UNIT SELECT 68 0 F Configure conductivity units 1 Select EC from the units menu 2 Select the...

Page 16: ...Device Address GPH MNU 0 10 50 0 F pH uS mV DWN ENT UP EXIT ALARMS CALIBRATION SETUP PUMPS MENU DWN ENT UP BACK TIME DATE UNITS LOG INTERVAL SETUP NEXT DWN ENT UP BACK COMM DEVICE ADD MFG INFO SETUP B...

Page 17: ...wtek Inc www agrowtek com WiFi AgrowNET Link Port TM WiFi Reset DigiHydro Water Monitor Controller TEMPERATURE CONDUCTIVITY pH O R P D O if equipped GrowControlTM POWER 15 24Vdc pH Probe o r p d o Pro...

Page 18: ...X8 GrowNET Hubs allow multiple GrowNET sensors relays and dosing pumps to be connected to a single LX1 or LX2 interface Indi vidually buffered full duplex ports for signal integrity Hubs can be daisy...

Page 19: ...value 700 To set the conductivity zero calibration write the value 0 Span Calibration Span or slope calibration corrects the slop of the sensor reading at a second point away from the zero cali bratio...

Page 20: ...libration Input Signed Int 16 bit W 41103 Sensor 4 Offset Cal Calibration Input Signed Int 16 bit W 41104 Sensor 1 Span Cal Calibration Input Signed Int 16 bit W 41201 Sensor 2 Span Cal Calibration In...

Page 21: ...treatments can be tried They are presented in the order of the severity of attack on the pH glass and may not improve and in some cases actually further deteriorate electrode performance DANGER Use pr...

Page 22: ...rranty Period In no event or circumstance shall Agrowtek Inc be liable to any third party or the claimant for damages in excess of the price paid for the product or for any loss of use inconvenience c...