FIVE. Calibration
5. Weight Calibration
★
Note: the self-calibration procedure should be run during initial installation, using or tested unbalanced weight is
suspected to be inaccurate, to ensure the accuracy of the balancing machine (Note that the 100g balance lead used for
self-calibration should be accurate, otherwise it is not correct about final result and directly affects the balance precision!)
5.1 Power on and turn on the machine's power switch.
5.2 Install a medium size (13 "-18") tire which can realize balance lead to be clipped on both inner and outer side of rim,
input rim data.
6. Distance ruler calibration
(
optional function
)
No automatic gauge, no calibration
required.
1
Press and hold
key, at the meanwhile press
key
display
>
2
Put down the hood, press
key to rotate the main shaft, wait until it
stops
display
>
3
Open the hood, rotate the rim to see outer indicator lights are all on, add
100 grams of balance lead at 12 o 'clock and lower the cover,
press
key to rotate main shaft, wait until it stops
display
>
4
Open the hood, rotate the rim to see inner indicator lights are all on, add
100 grams of balance lead at 12 o 'clock and lower the cover,
press
key to rotate the main shaft, wait until it stops.
display
>
Self-calibration is successful
!