9. Measuring using Measuring-Menus
Line
-----------------------------
P31
Blank line
P30
Text1
Comment Text 1
P37
Text2
Comment Text 2
P38
Text3
Menu Title U1
P39
Text4
Menu Title U2
P40
Text5
Menu Title U3
P41
Locking
LOCKING MODE: 5
P42
9.7.4 Programming via the Serial Interface:
Apart from the output errors mentioned above, only new commands, which are
not covered by the manual section 6, will be listed in the following.
Selecting the menu line xx:
ixx
Selection of menu u and function yy:
fu oyy
Entering the texts:
Text 1:
f5 $Text1
Text 2:
f6 $Text2
Text 3=MenuTitle U1:
f7 $Text3
Text 4=MenuTitle U2:
f8 $Text4
Text 5=MenuTitle U3:
f9 $Text5
Output of the texts 1 to 5:
f5 P20
Text1
etc.
Output of the menu configurations:
fu P20
Menu title of the menu u
U1:MenuTitle U1
In line 00: function yy
00:30
In line 01: function yy
01:39
....
02:16
03:24
04:30 ...
Output of all functions of the selected menu
(print format see above):
P20
Menu title
MeasCorrection
Meas. value average
00: +025.67 °C
RANGE: 00: Ntc
Blank line
LOCKING:0.
SETPOINT: 00: +0000.0øC
COMPENSATION:00:+0000.0øC
ATM.PRESSURE: +01013. mb
Input of the cross-section for the inp.channel in cm
2
:
Qxxxxx
(max. 32000, s. 9.4.9)
Input of the smoothing for the input channel:
f1 zxx
(s. 9.4.1)
Input of temperature compensation in 0.1 °C:
f1 gxxxxx
(f1 g00150=15.0°C)
Input of setpoint at input channel:
f2 gxxxxx
Setpoint adjustment (VM=4 Factor, VM≤3 Slope):
f2 C01
44 ALMEMO 2690-8