Measuring with one Measuring Point
Clear single value wich key:
<CLR>
Clear Max, Min and Avg. Values of all channels:
<CLRA>
Due to the continuous measurement the current measured value will be imme-
diately displayed after the clearing process. Furthermore, the peak values are
cleared at each start of a measurement, if the device has been correspond-
ingly configured (standard setting, s. 11.5.8).
10.2 Correction of the Meas. Value and Compensation
To achieve a maximum accuracy of measure-
ments it is possible to correct the zero point of
the sensores in all menus by pushing a button.
Additional correction functions are provided in
the ´User Menu´
U1 Meas. Correction
(selec-
tion s. 9.1). By entering a setpoint the correc-
tion value will be automatically calculated and
stored in the sensor connector. A correspond-
ing compensation is provided for sensors,
which are affected by the ambient temperature
or the atmospheric pressure.
10.2.1 Setting the Meas. Value to Zero
The user can zero the measured value at certain locations or at certain times
in order to check the deviation from this reference value. After selecting the
function Meas. Value (s. 9.4) in any menu, a help window shows you all possi-
bilities of correction of the measured value. The keys
<ZERO>
,
PROG
can be
used to store the displayed measured value as base value and, as a result, to
set it to zero.
Function select
Measuring Value
:
00:
23.4
°C
Function
Set Measuring Value to Zero:
<ZERO>
Carry out with key:
PROG
Measuring Value:
00:
00.0
°C º
Base Value:
Base:
23.4 °C
If the function is locked (s. 11.3.4), the base
value is not stored in the connector, but only
temporarily in the RAM until switch-off.
As long as the deviation from the base value is indicated (instead of
the actual measured value) the symbol
º
appears in the display.
The base value must be cleared in order to re-obtain the actual mea-
sured value (s. 11.3.6).
ALMEMO
®
2890-9
25
Sensor ist locked
-Zero setting temporarily
press key: PROG
-To cancel press key: ESC
C © REC COM l© ©l R01 * ´´´´´´µµµµ¶
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
01: 25.45 m/s
L840 Staurohr º
Locking mode:
3
Setpoint:
25.0 mls
5 Base:
-----
5 Factor:
0.6891
4 Zero corr:
0.7 mls
4 Slope corr:
-----
Temp. comp:
245.7 °C
Air. pressure:
1027 mb
START MANU M PRINT ESC